Authors

  1. Urbin, M. A. PhD
  2. Bailey, Ryan R. MS
  3. Lang, Catherine E. PhD

Abstract

Background and Purpose: Accelerometers have been used to capture real-world use of the paretic upper extremity in people with stroke. It may be possible to characterize different aspects of the recorded acceleration to gain insight about movement capabilities during task-specific behavior. These measures may be of value for guiding rehabilitation. We undertook a study to identify the acceleration characteristics that have a stable association with upper extremity function and sensitivity to within-participant fluctuations in function over multiple sessions of task-specific training.

 

Methods: Twenty-seven adults 6 months or more poststroke with upper extremity paresis participated. Signals from wrist-worn accelerometers were sampled at 30 Hz during 7 sessions of task-specific training. Paretic upper extremity function was evaluated with the Action Research Arm Test. We used Spearman correlations to examine within-session associations between acceleration metrics and Action Research Arm Test performance. A mixed model was used to determine which metrics were sensitive to within-participant fluctuations in upper extremity function across the 7 training sessions.

 

Results: Upper extremity function correlated with bilateral acceleration variability and use ratio during 5 and 6 sessions, respectively. Time accelerating between 76% and 100% of peak acceleration correlated with function in 6 sessions. Variability of the paretic upper extremity acceleration and the ratio of acceleration variability between upper extremities were associated with function during all 7 sessions. Variability in both the acceleration of the paretic upper extremity, and acceleration of the paretic and nonparetic extremities combined were sensitive to within-participant fluctuations in function across training sessions.

 

Discussion and Conclusions: Multiple features of the acceleration profile track with upper extremity function within and across sessions of task-specific training. It may be possible to monitor these features with accelerometers to index upper extremity function outside of clinical settings.

 

Video Abstract available for more insights from the authors (see Supplemental Digital Content 1, http://links.lww.com/JNPT/A91).