Keywords

Accelerometer, Physical activity, Sedentary, Smart ring, Validation

 

Authors

  1. Niela-Vilen, Hannakaisa PhD
  2. Azimi, Iman PhD
  3. Suorsa, Kristin PhD
  4. Sarhaddi, Fatemeh MSc
  5. Stenholm, Sari PhD
  6. Liljeberg, Pasi PhD
  7. Rahmani, Amir M. PhD
  8. Axelin, Anna PhD

Abstract

Smart rings, such as the Oura ring, might have potential in health monitoring. To be able to identify optimal devices for healthcare settings, validity studies are needed. The aim of this study was to compare the Oura smart ring estimates of steps and sedentary time with data from the ActiGraph accelerometer in a free-living context. A cross-sectional observational study design was used. A convenience sample of healthy adults (n = 42) participated in the study and wore an Oura smart ring and an ActiGraph accelerometer on the non-dominant hand continuously for 1 week. The participants completed a background questionnaire and filled out a daily log about their sleeping times and times when they did not wear the devices. The median age of the participants (n = 42) was 32 years (range, 18-46 years). In total, 191 (61% of the potential) days were compared. The Oura ring overestimated the step counts compared with the ActiGraph. The mean difference was 1416 steps (95% confidence interval, 739-2093 steps). Daily sedentary time was also overestimated by the ring; the mean difference was 17 minutes (95% confidence interval, -2 to 37 minutes). The use of the ring in nursing interventions needs to be considered.